Summary
Anna reinstates surface met communicationDetails
- FIXED: Surface met had not communicated with the rugged computer since 10 March 2023. This problem was resolved by power cycling the RS407 manually at the RADMet enclosure, changing the port number to COM7 as expected, and power cycling again. Troubleshooting doc here .
- FIXED: Filenames were non-standard since June 2022-Mar 2023 after an upgrade to Windows 10 (not sure how it was associated - that problem was fixed 18 Dec 2023 with a bulk rename on powershell).
- Skyriver had not been syncing Bodega surface met since April 2021. Resolved 22 Dec 2023.
- T/RH was showing NAN as soon as connectivity was possible. Uncertain on the rain gauge functionality. Multiple photographs taken.
- Lots of corrosion on MRR system (not antenna though). Will send photos to Metek/ATS for their advice.
- See future recommendations for follow-up.