Summary
System troubleshooting - TPS 50V PSDetails
Personnel: Douglas Alden
-
Radiometrics responded
- Check PS for signs of scorching or burned component smell. If no issues, reseat PS.
- They are going to look into cost of replacement unit.
- In lab, removed cover of TPS 50V PS. No signs of scorching or burned component smell.
-
Reseated PS and restarted RWP.
- TCP error 66 was listed in Birch Error log on WPC.
-
Rugged computer
- Did not boot. Cycled power and it turned on.
- Could not connect to internet. Cycled power on RV55 to fix issue.
- Restarted WPC and profiler started working.
- Cleared alarm on MegaRAID software the occurs every time WPC system starts up. It requires logging in to MegaRAID software and selecting silence alarm. This is a known issue.
- Replaced filters on RWP fans.
- Added new server ip ranges to RV55 Trusted IPs Inbound Friends and updated FileZilla Server Settings > Protocols Settings > Filters
- Upgrade FileZilla Server to 1.10.3. Testing lftp connection from skyriver and merced.