Photos

Summary

Data Download / Replaced TB with calibrated TB / Replaced Enclosure / Added Bird Spikes

Details

HB / LK 1:38PM

Data Download.
Replaced TB with calibrated TB. Replaced enclosure to give battery a longer life. Added bird spikes.

Old enclosure missing 1 screw and taped shut.

New SN: 71664-117
Old SN: 73682-817

Funnel slightly dirty/dusty. TB has filter that was cleaned and replaced in funnel. Data downloaded.

NEEDS NEW BULKHEAD CONNECTOR AND CAP FOR SAID CONNECTOR (for computer access to data logger). Chewing has occurred on the wiring that connects the TB to the DL4.

Battery at 3.57 and was replaced with a new battery. No sharpie on hand so left blank.

Logging after replacing enclosure and TB; clock set to correct date and time.

Taped gnawed cable. Funnel had thumb screws but replacement TB didn’t have thumb screw holes drilled into it so we used electrical tape to keep it securely fastened.