Summary
Data Download / Dismantled Wrecked Staff Gauge / Moved CameraDetails
Carly Ellis/Chad Hecht
Remembering the direct read cable broke in WY 2019
Pulled Solinst (Serial #0022075052) out around 2150 UTC
Communication error on first try tried to clear optic readers
Communication error on second try so switched ports on comp
Communication error on third try :(
Attempting to deploy backup Solinst (Serial #0022073079)
Communication error so switched to third port
Communication error so checked that COM port driver was up to date
Cleaned optical lenses again but still got comm error
Attempting to deploy backup Solinst (Serial #0022073385)
Another two comm failures
Attempting slightly older model Solinst (Serial #0101044518)
Communication error
Messed around more in Device Manager and saw a warning sign on solinst port
Don't think it was even in the list before
Found it right under Network adapters and above Portable Devices
Browsed to update driver and selected levelogger4_4 > solinst
Now comunicating w original Solinst (Serial #0022075052)
Solinst clock time is 5 min slow
Last reading was 8/27/2020
Downloaded data and synchronized clock
Scheduled start time for 2300 UTC
Chad dismantled wrecked staff gauge
Moved camera for better stream shots
2024-10-08 Note from Sarah Burnett: We don't have data from the levelogger between 8/27/2020 17:46 and 3/13/2021 23:00. Seems like the logger stopped working on 08/27/2020?