Summary

Data Download / Dismantled Wrecked Staff Gauge / Moved Camera

Details

Carly Ellis/Chad Hecht


Remembering the direct read cable broke in WY 2019


Pulled Solinst (Serial #0022075052) out around 2150 UTC

Communication error on first try tried to clear optic readers

Communication error on second try so switched ports on comp

Communication error on third try :(


Attempting to deploy backup Solinst (Serial #0022073079)

Communication error so switched to third port

Communication error so checked that COM port driver was up to date

Cleaned optical lenses again but still got comm error


Attempting to deploy backup Solinst (Serial #0022073385)

Another two comm failures


Attempting slightly older model Solinst (Serial #0101044518)

Communication error


Messed around more in Device Manager and saw a warning sign on solinst port

Don't think it was even in the list before

Found it right under Network adapters and above Portable Devices

Browsed to update driver and selected levelogger4_4 > solinst


Now comunicating w original Solinst (Serial #0022075052)

Solinst clock time is 5 min slow

Last reading was 8/27/2020

Downloaded data and synchronized clock

Scheduled start time for 2300 UTC


Chad dismantled wrecked staff gauge

Moved camera for better stream shots


2024-10-08 Note from Sarah Burnett: We don't have data from the levelogger between 8/27/2020 17:46 and 3/13/2021 23:00. Seems like the logger stopped working on 08/27/2020?