Summary
Maintenance Visit and EH&S Data DownloadDetails
Log file location: @skyriver.ucsd.edu/data/CW3E_data/CW3E_HCN_Archive/raw/019_greenhouse/Dep-Rec_LOG/greenhouse_2024-10-25.log
Greenhouse
Site 019
25 Oct 2024
Douglas Alden, Charlotte Piazza
Data download for EH&S and Site Maintenence Visit
On arrival
System operational?
T0 dangling out of housing - probably why temp was reading high, may need to nan some data
Pyranometer reading is about 20% low - was covered with some guano
Funnel looks a little clogged but not completely, water could probably flow okay
Battery Voltage: 14.3 V
Current: 3.4/1.5/1.3 mA
Clock OK? DL4 is +0m15s fast
Barometer offset
Paros: 1004.3
DL4: 1004.2
Current offset: +1.4
New offset: not changed
Old SD Card #: 096
New SD Card #: 036
Replaced old T/RH w/ Sensirion SHT-85
Old Anemometer: 49304
New Anemometer: 95681 Sep 2024
Old Pyranometer: PY110013 16 Jul 2019 -14.13
New Pyranometer: PY49878 06-Sep-2023 -11.07
Old Tipping Bucket: 31864-0103 May 2019
New Tipping Backet: 31855-0103 Sep 2024 ; tipper HCN-00X; plugged into TB1
Removed old 900 mhz antenna
Moved T/RH to beehive on the south side of the tower, removed the old beehive on the north side
Upgraded firmware to v2.03 Mt. Rainier
after restart
Battery Voltage: 14.3
Current: 2.7/1.3/1.3
Sensor Heights:
T0, TRH: 166 cm
Barometer: 427 cm
TB: 438 cm
T1: 706 cm
Pryanometer: 731 cm
Anemometer: 731+144 = 775 cm